Förster Resonance Energy Transfer (FRET) is used to detect molecular interactions within cells at the nanometre scale, enabling researchers to study protein–protein, protein–DNA, and other biomolecular interactions.
I trained as a physicist before I moved into biology, so when I started watching how my colleagues ran FRET experiments, I noticed that several controls were missing! This article explains why those controls are critical and how you include them in your FRET experiments.
Note: This article is based on Pragya’s FRET webinar, which you can watch here.
How FRET Works
FRET uses two fluorescent tags, one on each of the two molecules you’re studying, for example, two proteins. In FRET, these two tags are called the donor and the acceptor:
Choose a free resource to help you move forward
DOWNLOAD
Reveal More from Complex Imaging Data
EBOOK
Introduction to Computational Clearing
- The donor is the tag you shine light on
- The acceptor is the other tag. If it’s very close to the donor (within about 10 nm), the donor passes its energy straight to it, and the acceptor glows in its own color
True FRET only happens when the two tags are almost touching. So if the acceptor glows when you only shine light on the donor, it’s a sign that your two molecules are right next to each other. You can find out more about how FRET works here.
Your “FRET Channel” Doesn’t Only Contain FRET
In FRET imaging, you take several images of the same sample. The “FRET channel” is where we shine light meant to excite the donor and record only the acceptor’s color. If the acceptor glows, we assume the donor has transferred energy to it. This approach is called “sensitized emission”.
The trouble is that light can get into that channel in three different ways:
- True FRET. The donor passes its energy to an acceptor that is very close by (a few nanometers away), and the acceptor glows. This is the signal you want.
- Direct acceptor excitation. The laser you use to excite the donor also excites some of the acceptor directly, because the acceptor absorbs a wide range of colors. This is not FRET.
- Donor bleed-through. The donor gives off a wide range of colors, and some of that light leaks through the filter meant for the acceptor (real filters don’t have perfectly sharp cut-offs). This is not FRET either.
However, your camera can’t tell these three signals apart. It only measures how much light arrives, not where it came from. So a bright FRET channel could mean lots of FRET, or lots of leakage and very little FRET!
Leakage happens because each fluorophore absorbs and emits across a broad range of colors rather than a single wavelength. This broad range is what makes FRET possible, as the donor’s emission has to overlap with the acceptor’s absorption, but it is also what lets unwanted light into your measurement.
How much of each you get depends on your fluorophores, your filters, your camera or detector, how much donor and acceptor are in the sample, and your imaging settings. In some setups, this unwanted light is small. In others, it drowns out the real FRET signal. But you can’t know which until you measure it with controls.
Before You Start: Test Your Donors
But before you set up any controls, start by choosing your donor. The advice I give most often is to test more than one donor.
When I first set up my microscope for FRET, I tested several green donor tags with the same acceptor, using the same settings. How much light leaked into the FRET channel (β) varied a lot between donors, but on paper they all looked suitable.
Why Testing Donors Matters
Websites like FPbase tell you how well a donor and acceptor should work together on paper. But they can’t tell you how much of the donor’s light will leak into your FRET channel, since that depends on your microscope filters, camera, and settings.
If most of the light in your FRET channel is leakage, you have to subtract a lot to find the true FRET signal. The problem with this is that small errors in your leakage correction can become big errors in your FRET result. So I’d rather use a donor that gives a bit less FRET but leaks less than one that gives more FRET but also more leakage.
How to Do It
- Look at the colors first. In FPbase’s spectra viewer, plot your possible donors and acceptor together, then add the lasers and filters your microscope uses. You’ll see where the donor’s colors overlap the acceptor’s filter, i.e., where leakage comes from.
- Test two or three donors. For each donor tested, make a sample with only the donor and measure how much leaks into the FRET channel (for how to do that, see Part 1 below).
- Pick the donor that leaks least, not the one that looks best on paper.
- Do it on your own microscope. Some microscopes have a built-in FRET calibration routine. Mine didn’t, so I wrote my own in MATLAB. The calculation is simple and doesn’t need special equipment.
Because you need a donor-only control anyway (control 2 below), this adds very little work. Once you’ve chosen your donor, you’re ready to set up your controls.
Here Are the Four Controls You Need
To separate true FRET from leakage, you need four control samples. They do two different jobs:
- The first two samples are the calibrating controls. They tell you how much leakage to subtract.
- Samples three and four are the validating controls, or your positive and negative controls. They check that your molecules really are close together once leakage is subtracted. In my experience, these are the controls most often skipped.
| Control | What it tells you | Job |
|---|---|---|
| Acceptor-only sample | How much the acceptor glows under the donor’s light, with no donor there (α) | Calibrate |
| Donor-only sample | How much donor light leaks into the FRET channel (β) | Calibrate |
| Known-FRET sample | Positive control: do you see FRET when the two tags are definitely together? | Validate |
| Known-no-FRET sample | Negative control: do you see no FRET when the two tags are definitely apart? | Validate |
Controls 1 & 2: Measuring Leakage
As we saw earlier, two kinds of leakage add light to your FRET channel on top of true FRET signal:
- The first is direct acceptor excitation, where the light you shine to excite the donor also makes some of the acceptor glow
- The second is donor bleed-through, where some of the donor’s own glow slips through the filter meant for the acceptor
Controls 1 and 2 let you measure each one separately. Control 1 (acceptor only) shows how much direct acceptor excitation you get, and control 2 (donor only) shows how much bleed-through you get. You then subtract both values from your real sample.
How These Controls Work
- Control 1: Acceptor-only sample → α. This sample has the acceptor tag but no donor, so FRET can not happen here. Any light from this sample in the FRET channel must come from the donor’s light, making the acceptor glow directly. You should also see little or nothing in the donor image. If you do see signal there, the acceptor is leaking into that image too, and you’ll need to measure that before proceeding.
- Control 2: Donor-only sample → β. This sample has the donor tag but no acceptor, so FRET cannot happen in this sample. Any light in the FRET channel must be the donor’s own light leaking through the acceptor filter.
How to Calculate Leakage
Take three images of the above samples and subtract the background from each. From each image, measure how bright the signal is. This is its intensity, written as I.
The two letters after the I_ say which image your intensity measurement came from. The first letter is the tag you shine light on, and the second is the tag whose color you record (D = donor, A = acceptor):
- I_DD: excite the donor, collect donor light (donor channel)
- I_DA: excite the donor, collect acceptor light (FRET channel)
- I_AA: excite the acceptor, collect acceptor light (acceptor channel)
Each correction factor is just one intensity divided by another:
- α = I_DA / I_AA, measured on the acceptor-only sample
- β = I_DA / I_DD, measured on the donor-only sample
Then, for your real sample (the one with both tags), the corrected FRET signal is:
Three things to watch when doing these calculations:
- Check what you divide by. For α, you divide by the acceptor channel. For β, you divide by the donor channel. Getting these mixed up is the most common error I see
- Subtract background first. If you calculate the factors on images that still contain background, every corrected image after that will be wrong
- Check the symbols in papers. α and β are the convention I use here. Other papers use different letters, so check what each symbol means before you compare
Then Adjust for How Much Tag Each Cell Has
Subtracting the leakage leaves you with the true FRET signal (F_c). But F_c still depends on how much of each tag a cell happens to contain. A cell expressing lots of both tags gives a bigger F_c than a cell expressing a little, even if its proteins interact just as much. So the last step is to divide F_c by a measure of how much tag is there, much as you’d normalize a western blot band to a loading control. What you divide by changes what your final number means [1].
I use nFRET [2], which divides F_c by the square root of (donor signal × acceptor signal). It’s a number for comparing cells within one experiment, not a true FRET efficiency, and it only holds across days if your settings and calibration stay the same.
Controls 3 & 4: Positive and Negative Samples
As demonstrated, your calibrating controls tell you how much leakage to subtract. The remaining controls tell you if the remaining signal really means your molecules are close together.
Why Colocalization Isn’t FRET
On a standard widefield or confocal microscope, two labels up to about 200–300 nm apart can look as if they’re in the same spot (colocalized). But FRET happens over a much shorter range, roughly 1–10 nm, and it fades very quickly as the molecules move apart:
Here, E is the FRET efficiency (the share of the donor’s energy that reaches the acceptor), r is the distance between the two tags, and R₀ is the distance at which half the energy is transferred. In practice, FRET drops off very sharply: at twice that distance, under 2% of the energy gets through.
Think of two proteins in the same mitochondrion. Under the microscope, they look colocalized, but that doesn’t mean they’re interacting. While colocalization tells you two proteins are in the same compartment, FRET tells you they’re close enough to be touching.
I have images of vesicles labelled with the donor and acceptor tags that look as though they’re touching. By any reasonable measure, they’re colocalized, but they show no FRET at all. What told me was the negative- and positive-control samples.
Choosing Your Validating Controls
Your validating controls should differ in only one way: whether the molecules can get close together. Keep everything else as similar as possible. That way, any difference in FRET between the two can only come from how close the molecules are. How you keep the molecules apart in the negative control depends on your system:
- For my vesicles, I simply left out the molecular bridge that brings them together
- For a protein–protein interaction, it might be a mutant that can’t bind, or a version of your construct where the donor and acceptor tags are held too far apart for FRET
The positive and negative controls are the minimum. If you want to make a stronger claim, you may need extra checks, such as:
- Acceptor photobleaching. When FRET is happening, the donor passes some of its energy to the acceptor, meaning it glows less brightly than it would on its own. If you destroy the acceptor with strong light, the donor has nowhere to pass that energy and gets brighter. A brighter donor after bleaching is independent evidence of FRET.
- A titration. Gradually increase the amount of one partner. If the molecules really bind, the FRET signal should rise and then level off once all the binding partners are taken. If it keeps rising, the molecules may simply be crowded together rather than binding.
- A second negative control. Make another negative control that keeps the molecules apart in a different way. Your first negative control might show no FRET for an unrelated reason, such as a mutant protein that doesn’t fold properly, so a second one rules that out.
Without these controls, a FRET signal might hint that something is going on, but it can’t back up a firm, numbers-based claim that your molecules are close together.
Keep Your Calibration Valid
Your leakage measurements (α and β) only apply to the exact settings you measured them with: the same lasers and laser powers, filters, detector gain, and detection wavelength ranges. So if you change any of these, measure them again.
That means resisting a familiar reflex: turning up the laser or gain when a sample looks dim! Within one FRET experiment, it’s usually better to fix weak signals by improving the sample or the setup instead:
- Choose a donor that doesn’t fade quickly. It’s easier than correcting for fading (photobleaching) later. Antifade reagents and oxygen scavengers can help, but they change the sample’s chemistry, so use them sparingly.
- Reduce motion. The three images are taken one after another, so if anything moves in between, the pixels no longer line up, and the correction fails. An image splitter or two linked cameras can capture the donor and FRET channels at the same time (the acceptor channel still needs its own exposure, and the channels must be carefully aligned). For samples that move, thickening the medium (e.g., with agarose) helps, but it can also affect the biology.
- Control the environment. A temperature-controlled stage and a sealed imaging chamber are worth the setup time.
- Fix the sample if your question allows. Fixed samples are easier to image, but fixation can change a protein’s shape, how close proteins sit to each other, and how the tags behave.
Going Further
- Calculating a true FRET efficiency. The method above corrects for two sources of leakage. More complete methods, such as the one described by Gordon and colleagues [3], correct for four sources of leakage and calculate a true FRET efficiency. In my system, the extra two were small enough to ignore, but measure them in yours before you exclude them.
- FRET-FLIM. Instead of brightness, FLIM measures timing: how much faster the donor’s glow fades when a nearby acceptor drains its energy. Leakage doesn’t affect this readout, and it’s less affected by how much label is present. But how fast the donor fades also depends on its surroundings (such as pH), and you’ll need specialized detectors and software, plus your own set of controls.
Key Takeaways
In summary, before you commit to a donor fluorophore, prepare donor-only samples for two or three candidates and measure how much each leaks into your microscope’s FRET channel. Then choose the donor that leaks least. That one pilot experiment can decide whether your FRET signal is measurable at all.
And remember that your FRET channel holds more than FRET. To get a FRET signal you can trust:
- Calibrate with donor-only and acceptor-only samples to measure and subtract the unwanted light.
- Validate with known-FRET and known-no-FRET samples to confirm your signal really means the molecules are close together.
- Keep your settings fixed or recalibrate when they change.
Setting all this up took me a few months at first, and it felt like a detour. It wasn’t. Once your system is calibrated, the workflow becomes routine.
Good luck with your next FRET experiment: keep calm and don’t fret!
References
- Zal T, Gascoigne NRJ. Photobleaching-corrected FRET efficiency imaging of live cells. Biophys J. 2004 Jun;86(6):3923-39. doi:10.1529/biophysj.103.022087
- Xia Z, Liu Y. Reliable and global measurement of fluorescence resonance energy transfer using fluorescence microscopes. Biophys J. 2001 Oct;81(4):2395-402. doi:10.1016/S0006-3495(01)75886-9
- Gordon GW, Berry G, Liang XH, Levine B, Herman B. Quantitative fluorescence resonance energy transfer measurements using fluorescence microscopy. Biophys J. 1998 May;74(5):2702-13. doi:10.1016/S0006-3495(98)77976-7
You made it to the end—nice work! If you’re the kind of scientist who likes figuring things out without wasting half a day on trial and error, you’ll love our newsletter. Get 3 quick reads a week, packed with hard-won lab wisdom. Join FREE here.