Search below to delve into the Bitesize Bio archive. Here, you’ll find over two decades of the best articles, live events, podcasts, and resources, created by real experts and passionate mentors, to help you improve as a bioscientist. Whether you’re looking to learn something new or dig deep into a topic, you’ll find trustworthy, human-crafted content that’s ready to inspire and guide you.
When performing an un-established analytical method, one needs to statistically validate it. Failing to do so, you might end up with erroneous results and waste time and resources. Limit of detection (LOD) and limit of quantitation (LOQ) are the two crucial parameters used to validate analytical methods. This article will describe what LOD and LOQ are, their importance in method validation, and finally a quick method to calculate these values using Excel.
What Are LOD and LOQ?
LOD and LOQ mean exactly what these terms suggest. They are the minimum amount of compound (analyte) that can be reliably detected and quantified, respectively, using an analytical method/instrument. [1]
When measuring the concentration of an analyte, the analyte must produce a signal greater than the analytical noise, i.e., the signal obtained in the absence of analyte. [1] If the analyte concentration is too low, the analyte signal cannot be distinguished from analytical noise. For this reason, a limit is set for the detection of the analyte (LOD), which is higher than the signals that fall in the analytical noise zone. This ensures that the signal is indeed due the analyte, and not due the analytical noise.
The Importance of Computing LOD and LOQ
In Validating a New Analytical Method
New analytical techniques can have a lot of advantages over tried-and-true methods. For example, HPLTC can be more cost effective and sensitive than the more traditional HPLC. If the concentration of the analyte you plan to detect using a certain analytical method that you developed is below the LOD and/or LOQ, then the signal you detect cannot be trusted. To ensure that your data are accurate and reliable, it is important to compute the LOD and LOQ for any new analytical technique.
Choose a free resource to help you move forward
CHEAT SHEET
Lab Math Cheat Sheet
There are some lab math calculations that almost all molecular biologists need to perform, and often on a regular basis. Make sure you never mess up a solution or miscalculate your DNA concentration again, with our handy lab math cheat sheet. Print it out and stick it on your bench.
Save time and hassle when making up your solutions or calculating your DNA concentration, thanks this lab math calculator spreadsheet. Preloaded with common calculations for molecular biologists, the spreadsheet allows you to easily and quickly get the answers you need.
It is also wise to compute LOD and LOQ for an already established method. Although you might try your best to avoid incorporating manual errors, there is still a good chance of that happening. For instance, pipetting errors can lead to variations in response when performing serial dilutions to prepare a standard curve. Furthermore, the sensitivity of every analytical technique differs with the instrument. Therefore, the standard curve you have freshly plotted requires statistical validation to ensure the responses of your test samples are within the range of detection/quantitation.
Calculating the LOD and LOQ Using MS Excel
Among the methods used to validate analytical methods, such as serial dilutions, computing LOD and LOQ via a calibration curve in MS Excel is more accurate and reliable. The serial dilution method defines the LOD as the concentration of analyte that gives you a response 95% of the time across a dilution series. Although the dilution subsequent to the assumed LOD (say ‘x’) might not give a response 95% of the time, there is a chance that a concentration between the assumed LOD and ‘x’ might. Thus, in the case of the serial dilution method, there is no exact figure that you can accurately/statistically term as the LOD. However, it may be used to verify the LOD that has been determined via other methods. [2]
On the other hand, the calibration curve method using MS Excel is based on the standard deviation (SD) of response (\(\sigma\)) and slope (s) of the calibration curve, [3] thus making it more accurate. The SD (\(\sigma\)) can be calculated for the Y-intercept and used to calculate the LOD and LOQ for a standard curve using the regression function in MS Excel (I use version 2013; however, you can use any version and it’ll work just fine) as follows.
Step 1: Plot a Standard Curve
Construct a standard curve by plotting the concentration of analyte on the X-axis and the response on the Y-axis. Assuming the response is directly proportional to the analyte concentration, you should plot a linear curve.
Step 2: Carry Out a Regression Analysis Using the Regression Function
You can find the regression function under Data>Data Analysis>Regression. Under the regression function, select the data for the X and Y variables.
The regression function yields three outputs: regression statistics, ANOVA, and coefficients. Regression statistics tell you how well the regression equation fits the data, ANOVA gives you the level of variability within your regression model, and the coefficients (the most useful part) include the slope of the curve and the SD of the Y intercept. You now have a new Excel sheet with the data required to compute the LOD and LOQ.
Step 3: Use the \(\sigma\) and s values in the following formulae to compute the LOD and LOQ
LOD = [3.3*(\(\sigma\)/s)]
LOQ = [10*(\(\sigma\)/s)]
The coefficients 3.3 and 10 are called expansion factors and are obtained assuming a 95% confidence level. [4]
By computing the LOD and LOQ for your standard curve using this quick and simple method, you can rely on your data. Try this method to validate your analytical technique and let us know in the comments if you have excelled with Excel.
You made it to the end—nice work! If you’re the kind of scientist who likes figuring things out without wasting half a day on trial and error, you’ll love our newsletter. Get 3 quick reads a week, packed with hard-won lab wisdom. Join FREE here.
Have you ever wondered what it would be like if someone helped you step-by-step through your job search? A good recruiter does exactly that! Recruiters provide value to job-seekers by reviewing resumes, finding jobs that may be a good fit, and providing interview tips. But how does that process work? In this article we’ll cover…
So now you’re convinced that R is the language for you, you’ve downloaded R-Studio (from https://www.rstudio.com/) and opened it, and. . .what the hell do you do now? Great question! I always find it easiest to learn by doing something, rather than just by seeing a list of possibilities, so here I’ll walk you through…
In my first article on this topic we delved into what miRNAs are, how they are generated, and their function. Now, we are going to discuss how to identify miRNAs and their targets. Why Do You Want to Look at Something So Small Anyhow? miRNAs play a crucial role in most physiological processes. It’s not…
Stuff moves. It is useful to study how stuff moves, because motion analysis can tell us a lot about the object that is moving. For example, we can learn if an object’s motion is aimless, diffusive wandering, or directed towards some goal, free to explore the available environment, or restricted to a confined space. Studying…
Last time we brushed up on cell referencing and constructing formulae to use Excel for some basic and some slightly more advanced calculations. This time we’re going to move on to using some built-in Excel functions and go through how to apply all that we’ve done so far to a worked example that is relevant…
1-2-3 Newsletter
Get help with everything* lab-related.
*Well, everything except the washing up. That’s still on you.
10 Things Every Molecular Biologist Should Know
The eBook with top tips from our Researcher community.