Ever done a multicolour flow cytometry experiment, run all your controls, done your compensation and then started to analyse your data and realised that you can’t work out where to put your gates?
To err is human…
When acquiring data on a cytometer, there can be measurement errors due to counting statistics, errors in processing the data and inaccuracies of log amplifiers. When data is compensated these errors are revealed and there will be spreading of your data. The more photons that are counted (the cells get brighter) so the measurement error increases. This causes all the channels to spread when correctly compensated but we notice it more on the brighter populations.
You should be aware of these errors and in order to minimise the problems of spreading of your data, you can perhaps design a better experiment, or avoid channels that will have large spreading of data. However, make sure you have the correct controls to make sense of your data and interpret it correctly.
Fluorescence minus one controls
Fluorescence Minus One (FMO) controls are now commonly used in multicolour experiments and they must be run with your experiment. These controls are cells (same cell type as your sample) stained with all the fluorochromes minus one fluorochrome. This allows you to take into account any spread of fluorochrome(s) into the unlabelled channel and allow you to put your gate in the correct place. They are essentially your gating controls for your experiment.
Example Data:This data shows FITC and PE staining from the panel and displays the accurate identification of the CD4 (PE+) T cell population using the FMO control. The plots show that even though you can’t see the PE-Cy5 and PE-Cy7 data, the compensation between the channels reveals the spread of data in the PE channel. Therefore, in order to accurately gate the PE+ cells the gate should be placed at the top of the FMO staining and not the unstained control.
Are they really necessary?
Yes, they are a great control and you will find more and more reviewers asking for them. However, you might not need them for all channels. FMO controls are only crucial when accurate identification between a positive and negative population is vital, e.g. when you have dimly expressed antigens or a continuous distribution of staining. Look at your panel, and see which channels need an FMO. It might be worth trying FMO controls for all samples once and then deciding which ones you want to look every time you do your experiment.
Golden rules : Always use controls in your experiments and it is better to run them at the time of your experiment than to have to repeat the whole experiment because of reviewers asking for them.Data shown from Mario Roederer, NIH Vaccine Centre.
The flow cytometer that we have all grown to know and love may have only come into its own in the 1990’s, but who would have known that the first cell sorter was invented as early as the 1950’s? With the recent death of one of the key developers of fluorescence activated cell sorting (FACS),…
Flow cytometers and cell sorters were designed with blood cells in mind. This means that commercial cell sorters are optimized for sorting cells typically smaller than about 20 µm in diameter. However, it turns out that many cell types, including those of mammals, are larger than 20 µm. So what are your options if you…
Like the legendary fight between boxers Bowen and Burke in 1893, the cell cycle in some cells goes on and on..round 1, round 2, round 3, round 4…before the final bell is rung. Nucleotide analogs, like BrdU or EdU, are great for examining 1-2 cell cycle division(s). In many studies though, for example the proliferation…
Flow cytometry is fast evolving from a method only revered by immunologists, to one used by nearly every biological specialty. It’s pretty much my favorite tool. Unfortunately, as with most lab techniques, much of flow cytometry is taught on the job without a lot of standards. And too often bad habits are passed along like…
If there are a million cells of interest in your sample and you pass them through a sorter, you might expect to get a million cells back. But you don’t – and here is why. Hardware aborts A hardware abort occurs when an instrument can’t process the information about an event because it is still…
While many scientists are methodical and precise, some of us like to live on the edge. Read a protocol all the way through? No thanks, I’ll take my chances and guess what concentration of HCl I should use. Label my tubes with the correct content? Puh-lease – it’s much more exciting deducing which is which…
10 Things Every Molecular Biologist Should Know
The eBook with top tips from our Researcher community.