Skip to content

Running You Through HPLC: How Does it Work?

Posted in: Analytical Chemistry and Chromatography Techniques
Running You Through HPLC: How Does it Work?

Your advisor tells you that he wants you to use HPLC to analyze your compound. You know you’ve heard of this technique before, but you can’t remember what HPLC stands for, let alone how to go about doing it! We’ve all been there, and I bet you wish you had paid more attention in that lecture!

Fear not – in this series of articles I will remind you about the power of the HPLC column  🙂

In this first article, I will take you through the principle behind HPLC and remind you of its uses – you will be ready for the lab in no time!

How does HPLC work?

High-performance liquid chromatography, or HPLC, is a long name for a powerful technique based on the simple fact that individual compounds behave differently in water.

HPLC separates and purifies compounds according to their polarity, or their tendency to like or dislike water. To put polarity into context, consider that oil is an apolar liquid that doesn’t mix with water. Ethanol, on the other hand, is polar and as many of you know, mixes very well with water (Vodka and coke anyone??).

I have tried to simplify the whole process in Figure 1 below, but first let’s look at the main components involved in HPLC. Apologies in advance for some unavoidable jargon!

The components

The HPLC column

This is also known as the stationary phase. This is the workhorse of the HPLC machine, is made from one of a variety of substances (often silica) and is very compact in nature. The silica particles are functionalized by long carbon chains. The carbon chains are apolar and therefore the longer the chain, the more apolar the column becomes. Columns containing 18-carbon chains are commonly used and are known as C18 columns.

The HPLC sample

Sample types vary greatly depending on the field and the type of compounds in question. HPLC can be used to analyze compounds in biological specimens (urine, blood, saliva, and muscle), environmental samples, medicinal chemistry (drugs), and microbiology (toxins produced by fungi and bacteria).

Injection of sample

Samples are injected into the HPLC column. This used to be carried out manually, meaning that some poor soul had to sit by the HPLC machine for hours on end injecting each sample with a syringe, sometimes all night long!

Luckily, newer models have an automatic injector, reducing the manual input and allowing higher throughput. Modern machines are equipped with software allowing the user to input a list of samples, how much and in what order they should be injected. So you can enjoy your lunch while your HPLC runs itself!

The mobile phase

This is really just a mixture of water and an organic solvent (usually acetonitrile or methanol). The mobile phase gets it name because it moves  through the column and at the same time elutes (or flushes out) the compounds from the column.

Compounds are often eluted along a concentration gradient. If you’re anything like me, concentration and gradient are two words you hate to see coming together in a sentence! It just means that the percentage of water in the mobile phase decreases over time, while the percentage of the apolar solvent increases simultaneously. This means that the mobile phase gradually becomes more apolar. Don’t worry too much about gradients for now, as they will appear again in a follow-up article.

The HPLC run

HPLC can be performed in a number of modes. The most commonly used method is known as reversed-phase (RP-HPLC) and this is what I describe here. In this mode, compounds are separated starting with the most polar, and ending with the apolar compounds. For all modes, a high-powered pump moves the sample and the mobile phase through the column. A  typical run can take between 10-60 minutes.

The principle behind HPLC – a closer look

Now that you have an idea of the components involved, let’s move on to the principle in a little more detail.

I mentioned above that HPLC separates compounds on the basis of polarity. But how does this actually work? English please! As the gradient kicks in, the solvent concentration increases while the water concentration decreases. This makes the mobile phase more and more apolar. Compounds contained in the sample will stick to the carbon chains in the column, with the most apolar compounds sticking the strongest, and the most polar compounds sticking weakly.

Figure 1 shows what happens to a sample containing a mixture of compounds after injection into the column. The compounds bind to the column and are flushed out at different times, depending on whether they are more likely to stick to the column or the mobile phase as it is pumped through. The time that each compound elutes (or flushes out) from the column is known as that compound’s retention time (Rf).

Figure 1: The principle behind HPLC

HPLC figure 1
Figure 1: Compounds of differing polarities (indicated as darkening shades of blue) are injected into the HPLC column (entire cylinder). The mobile phase is pumped through the column, and the addition of solvent along a concentration gradient (shown as a black dotted line) continuously decreases the overall polarity of the mobile phase (Y-axis). Compounds are able to stick to either the column or the mobile phase, depending on how polar they are. Compounds will eventually stick to the mobile phase when their polarity matches that of the mobile phase. They will then dissociate from the column and will be eluted at a particular time (X-axis) during the run. This time is known as the Rf for that compound.

Understanding the output

The output or results of an HPLC run is usually viewed as a chromatogram (Figure 2). This is a horizontal series of peaks representing compounds eluted from the column with different Rf values. Modern HPLC equipment is often coupled to a diode array detector (DAD), allowing the user to look at the resulting chromatogram of separated compounds in wavelengths from 190 nm to 900 nm. If the compounds under investigation are known, the user can choose to look only at 1 or a few selected wavelengths. For instance, cocaine can be observed at 254 nm.

Figure 2: A typical HPLC chromatogram

HPLC fig 2

Figure 2: This chromatogram shows the separation of compounds from a chemical reaction, and the chromatogram is viewed at 254 nm. Two main peaks occur at 8.20 and 9 minutes, representing two compounds with these retention times. The number of absorbance units (AU) is shown on the Y-axis while the time of the run is shown on the X-axis.

Applications

Within biology and medicine, HPLC is often used as an analytical tool to assay biological and environmental samples for the presence or absence of known compounds (for example metabolites, drugs, toxins, pesticides), and can assist in the identification of unknown compounds.

Within chemistry, however, HPLC is routinely used to monitor chemical reactions, as well as to determine the purity of the products. In addition, the process of HPLC can be modified to preparative HPLC, whereby compounds of interest can be purified for further use.

HPLC can come across as being very complicated to begin with, but rest assured, just like most other lab techniques, it makes a lot more sense when you actually do it.

Stay tuned for my follow-up articles in the coming weeks, in which I will go through some pointers on how you can get the best from your HPLC run depending on your research goal, as well as discuss objectively how HPLC can be used within your research.

Are there any other aspects of this technique you would like covered in more detail? If so, we would love to hear from you!

Happy HPLC-ing 🙂

Share this to your network:

6 Comments

  1. mbionwu Ugochukwu on June 17, 2019 at 11:34 pm

    how many samples can HPlC analyse at a given time

  2. jess on May 2, 2018 at 4:08 pm

    I thought this was a great article to explain a complicated process to someone with not much science. Very very good.

  3. Habibullah on January 24, 2018 at 6:57 am

    I want to determine pesticides residues and fungal toxin in agriculture food stuff what i should do get the result?

  4. Ken on September 19, 2017 at 6:25 pm

    How is the HPLC used to measure the levels of MDA in serum and tissue of oxidatively stressed rat model… Thank you!

  5. Mudassir on April 19, 2016 at 11:39 am

    I want to separate components present in leaves of plants. I was thicking which HPLC column can be suitable but after reading your article i got the point
    Thank you
    Regards,

    Mudassir

Leave a Comment

You must be logged in to post a comment.

This site uses Akismet to reduce spam. Learn how your comment data is processed.

Scroll To Top