How to Make Accurate Stock Solutions

Posted in: Basic Lab Skills and Know-how
How to Make Accurate Stock Solutions

Is it just me who thinks that biology researchers can be a bit sloppy when it comes to making and dispensing reagents accurately? Analytical chemists would laugh at our faith measuring cylinders for anything other than fairly rough work like making up media.

When it comes to more accurate work, like making up assay standards, measuring cylinders just just don’t cut it anymore. Here are a few ways to ensure that your stock solutions/standards are as accurate as possible, allowing you to get your experiments off to the best possible start.


1. Make up solutions and standards using volumetric flasks.

For accurate and reproducible stock solutions your tool of choice should be a volumetric flask. Volumetric flasks are far more accurate than measuring cylinders and pipettes, especially if you use class A flasks, which are manufactured to extremely stringent standards. Volumetric flasks are available in various sizes from 1 mL upwards and you can find a guide to using them here.

2. Use the correct balance, correctly.

Obviously there’s no point in using a highly accurate volumetric flask to measure the solvent volume, only to weigh out the solid using a balance that is too coarse for your needs. Make sure you use a balance that is made for weighing a mass in the range you require, and that it is calibrated and on a level surface.

For small masses, it is often quite difficult to weigh out exactly the amount you require. But normally , solutions don’t have to be made up to an exact concentration – you just need to know their concentration (whatever it is) exactly.

A good approach is to weigh out the reagent as close to the target weight as possible then, note down the weighed mass and make up the solution in the volumetric flask. The actual concentration of the solution you made up can then be calculated exactly..

3. Make reagents in large batches where possible

Another way to ensure that stock solutions are reproducible from experiment to experiment is to make them up in large batches where possible. This not only means that you are working with the same solution in each experiment, but also that you are using a larger volume when making up the solution, which should help with the accuracy.

4. Take into account the strength of reagents you are using.

When making up stock solutions, the strength of the reagent is often overlooked. If the reagent is less than 100% strength i.e. less than 100% of the mass is the actual reagent (the remainder being impurities), it will normally be stated by the manufacturer on the container and you should take this into account when calculating how much of it to use.

Say, for example, you need to weigh out 20g of a reagent to make a 1M solution, but the reagent is only manufactured to 97% percent strength. Only 97% (19.4g) of that 20g will be your reagent so you actually have to add 20.61g ([100/97]*20) of the solid to give you 20g of reagent… and 0.61g of the impurit(y/ies).

Do you agree that many biology researchers could/should do things more accuracy? Shout out your opinion in the comments.

Photo: Dublin Ninja

Share this to your network:

10 Comments

  1. Desmond on May 20, 2009 at 2:03 am

    Hi just wondering how do we keep our frozen items correct during stocktaking, almost every month we do facing different in the weight. example April balance 10 ton than on May balance 5 ton which this stock hardly move the most we move is only 1-2 ton per month? please advise thanks

    • Nick on May 20, 2009 at 8:30 am

      Desmond – look out for an article coming up on that soon

  2. Rob on July 17, 2008 at 6:11 am

    Another one from me. That’s a good point about the impurities Nick (never thought about that) but do any researchers actually take into account the impurities? If they don’t (and i’m assuming most don’t – i never have!), this would mean that slightly off concentration solutions are used by everyone to record their measurements and so for the sake of consistency we should all do the same and ignore the impurities. Any validity to that?

  3. Nick on July 17, 2008 at 3:04 am

    Good point Rob. My approach to that is normally to use the solvent to wash the weigh boat out into the beaker I am making the solution up in. But weighing the weight boat is also a good idea – I had never thought of that.

  4. Rob on July 17, 2008 at 2:54 am

    With regards to point 2, another good idea is to weigh the weigh boat after adding the solute to the flask. Then you know exactly how much went in there. There always seems to be something left over on the weigh boat after you’ve added the solute.